Stress Test System

The stress test system is a high-performance digital device capable of analyzing a wide range of parameters, including 12-lead ECG, ST and QT intervals, and more. Its software can control various medical treadmills and ergometers. The software continuously records the ECG along with heart rate variability (HRV) and automatically manages treadmill's operation.

To ensure a comprehensive assessment of heart health, is crucial to evaluate cardiac function both at rest and during vigorous exercise, as many arrhythmias intensify with physical activity.
The cardiac stress test is designed for this purpose. During the test, the patient exercises on a treadmill, which gradually increases to elevate heart rate. This allows the physician to determine the presence of coronary artery disease or arrhythmias, or to identify other potential issues.
The treadmill features an adjustable incline that gradually increases as the test progresses. Continuous blood pressure and heart rate monitoring to ensure that any abnormalities are detected and documented. An electrocardiogram (ECG) is used simultaneously to monitor the heart’s electrical activity.
For accurate readings, the cardiac stress test technician applies adhesive ECG Electrodes to the patient’s legs, arms, and chest. These Electrodes transmit data via cables to a computer for recording. Blood pressure is measured using a cuff placed on the patient’s arm.

Features

  • Efficient storage and management of all patient information
  • Detailed reporting and printing capabilities for complete stress test result, including stage-by-stage analysis and final summary
  • Support for established protocols, including BRUCE
  • Flexible to incorporate new protocols tailored to specific physician and patient requirements
  • Accurate measurement of standard parameters
  • Compatible with a range of medical treadmills

Treadmill specification

  • Maximum user weight capacity: 150 kg
  • Motor power: 4 HP

Ordering code

ProductOrdering code
Stex 9100T Treadmill6086
Wireless 12-Channel Stress Test Device5029

Technical specification

Patient Protectionfloating common (IEC-60601, CF type)
IsolationDC/DC converter, USB isolator
Shock ProtectionClass II.
Defibrillator Protection5 kV (max. 1 kV/µs)
Resolution0.05 µV/bit (24-bit A/D)
Sample Rate2000 Hz
Input Impedance>10 Mohm
CMRR>120 dB
Filters524 Hz low pass (hardware, constant)
50, 100, 200 Hz (software, switchable)
60, 120 Hz (software, switchable)
35 Hz (software tremor filter, switchable)
Time Constant 3.2 sec/ 1.6 sec
Linearity Error <0,5%, < 1/2 LSB
Polarization Voltage400 mV
Leads12 Standard + 3 Frank